GENERAL OPERATION
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
2-11
Print System Log
The system log prints the instrument serial number and configuration options along with
messages occurring in the HFA II. This feature is designed to be used by Carl Zeiss Meditec
field software representatives. Should you experience a problem with your instrument, it is a
good idea to print the system log before calling Carl Zeiss Meditec customer service.
Save/Transmit Option
This option allows the user to change the SAVE button functionality on the End of Test screen.
Choices are Transmit Only, Save Only, or Save and Transmit. Transmit allows the user to send
(transmit) data via a serial interface cable to an outside computer system at the end of every
test. Additional transfer methods may be available.
Printer
This option allows you to designate the printer type to be used with your instrument. The
choices are HP LaserJet and Printrex. The HFA II supports the following HP LaserJet Models:
1100 SE, 1200 and 3200 SE printers. The Lexmark Optra E312L printer is also supported.
These specific models of the printers were tested for functionality and leakage current
requirements. Other printers may also work with the HFA II - i series instrument.
Note: It is the owner’s responsibility to ensure that any printer used in a medical
environment meets the appropriate medical directives and International Safety Standards.
VA (Visual Acuity) Format
Select 20/20 Snellen, 6/6 Metric, or 1.0 Decimal as the visual acuity format used when entering
patient data.
Auto Pupil
(model 750i only)
If Auto Pupil is set to ON, the HFA II will automatically take a measurement of the patient’s
pupil diameter and enter the finding on the Patient Data 2 screen. An asterisk (*) is added
whenever the measurement was made automatically. The measured pupil size will also appear
on the display screen. Manual pupil measurement input is displayed without an asterisk. The
pupil diameter will also appear on the printout. This feature only works if Gaze Tracking has
been successfully initialized. For more information on Gaze Tracking, see Section 5: “Gaze
Tracking”.
Personalized I.D.
This allows you to customize hard copy printouts with 5 lines of text (e.g. practice name,
address, and telephone number).
Alter Main Menu
This allows you to customize the Main Menu screen by adding test buttons which normally are
only accessible through the test library, by deleting test buttons which are not often used, or by
altering test buttons to power-up with your preferred testing parameters. Additional text may be
added to further describe the parameters or usage of customized buttons. See “Altering the
Main Menu Screen” for details.
Additional Setup
This allows you to use additional System Setup functions found on the Additional Setup screen.
2.htp
3/14/03, 1:51 PM
11
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......