CARE AND CLEANING
12-5
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
3
While standing in front of the instrument, slide the
wishbone-shaped plate that holds the bulb
assembly in place away from you.
4
Remove the bulb assembly. Insert the replacement
bulb into the housing. Note the notch in the base of
the bulb housing lines up with the pin to the right of the
assembly. Do NOT touch the glass part of the bulb with
your fingers. If your finger touches the bulb, wipe the
bulb with a soft cloth.
5
Slide plate back into position, tighten screw and
replace connector cable.
6
Replace the top access panel.
12.htp
3/14/03, 2:45 PM
5
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......