7-28
STATPAC ANALYSIS & PRINTING RESULTS
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Printing test results stored
You can obtain printouts at any time convenient to you, if you store the results on disk. You
on disk
begin the process by selecting the
PRINT FUNCTIONS
icon from any screen where the
FUNCTIONS
icon is active. First you choose the Source and press PROCEED. This will retrieve
the File Directory. After selecting the test or tests from the directory, the program will either
start printing or present additional options, depending on the type of tests (screening or
threshold) selected. Refer to the chart below for more details.
If you select:
The print program:
1.
One screening test
Starts printing immediately. The print format
(right or left eye)
depends on the screening strategy used.
2.
Two screening tests for the
Presents the printout selection menu.
same patient, one for right eye
Screening tests are printed one per page or both
and one for left eye
on one page.
3.
Two or more tests for
Starts printing immediately. Threshold tests are
different patients
printed using the Single Field Analysis format,
where appropriate; otherwise the printout is the
Three-in-One format. Screening tests are printed
one page per eye.
4.
One threshold test
Presents the printout selection menu. If you
select Single Field Analysis or the Three-in-One
format, the program starts printing immediately.
If you select Overview, Change Analysis, or
Glaucoma Change Probability, the HFA II
presents the file directory with all eligible tests
selected (with a check mark). You may remove
tests from the list (by removing the check mark).
Press PROCEED when ready to print.
5.
Two threshold tests for the same
Same as #4.
patient, one for right eye and
one for left eye
Note: If you want to combine 24-2 and 30-2 results on one printout, STATPAC will
analyze only the Central 24 degrees. If you want an analysis of the Central 30 degrees, do
not combine 24-2 and 30-2 results for these printouts. The 10-2 test cannot be used in
conjunction with 24-2 or 30-2 tests, and is not available in the Glaucoma Change
Probability format.
PRINTING
PREVIOUSLY SAVED
TEST RESULTS
7.htp
3/14/03, 3:17 PM
28
Содержание HUMPHREY 720i
Страница 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Страница 4: ......
Страница 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Страница 327: ......
Страница 328: ......