Appendix
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
A-20
This section summarizes the data collection methodology (also described in the previous
section) and further describes the analysis and characteristics of the normative limits for
Ganglion Cell Analysis parameters.
cáÖìêÉ=^JNN==qóéáÅ~ä=d~åÖäáçå=`Éää=^å~äóëáë=ëÜçïáåÖ=`çãé~êáëçå=íç=kçêã~ä=iáãáíë
Methods
Data Collection
Data collection details are described in
“RNFL and Macula Normative Databases:
. In summary, 282 subjects were qualified as normal subjects,
enrolled in the NDB study, and included for the Macular normative database. Data from
the same scans was analyzed and evaluated to determine the normal range of Ganglion
Cell Analysis parameters. The normative database does not have subjects with refractive
errors outside the –12.00 D to +8.00 D range. Therefore, the analysis below should be
applied with caution to subjects with refractive errors outside the –12.00 D to +8.00 D
range.
Scan Selection Criteria
The scans were reviewed for image quality as part of the original study. One best quality
scan for each scan type was chosen for each subject per eye. Scans having the following
characteristics were excluded from the normative database:
• Signal Strength of 5 or lower.
• Large eye motion during image acquisition, resulting in a saccade that was within the
central 80% of the scan area.
• Area of data loss greater than 10% at the edge of the scan area.
• Presence of floater(s) obscuring macular area on Macular cube scans or measurement
circle on Optic Disc cube scans.
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...