Advanced Export
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
11-16
Advanced Export
Click on the Advanced Export button
on the ONH and RNFL OU Analysis screen, to
export and store ILM-RNFL thickness maps as .DAT files and .txt files of Neuro-retinal Rim
Thickness and RNFL Thickness profiles to a user-selectable directory
Dat Files of ILM–RNFL Thickness Maps
NOTE: DAT files are plain text files that can be read from an Excel or Matlab application.
Each comma–delimited value is the thickness of an A-scan. Each row represents a fast
B-scan. Each column represents a slow B-scan, as shown in the figure below. The first row
represents the topmost fast B-scan and the first column represents the leftmost slow
B-scan.
Txt Profiles of Neuro–retinal Rim Thickness and RNFL Thickness
The profiles of Neuro–retinal Rim Thickness and RNFL Thickness are saved to .txt files by
clicking on the Advanced Export button on the ONH and RNFL OU Analysis screen. The
files are plain text files.
The neuro–retinal rim thickness values at 180 points, each point representing 2 degrees,
are saved. Furthermore, the RNFL thickness values at 256 points, each point representing
1.41 degrees, are saved in the files.
For both profiles, the thickness data points start from temporal, to superior, nasal, inferior
and finally back to temporal — the same direction as on the screen. Patient and exam
information is also saved in these files.
Move Scan Data (Native Mode Only)
Occasionally, in Native Mode, a scan is acquired and saved in the wrong person’s file.
Scans that are incorrectly stored can be moved to a different file by using the Move Scan
feature. To access the Move Scan dialog:
1. Open the file with the incorrect or misplaced scan and choose an analysis to display
the data. For example, for a Macula Cube scan, choose Macular Thickness analysis.
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...