CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Analysis Related Report Options
10-15
statistically significant if the range of rates covered by the 95% confidence limits exclude
zero.
It may be more useful to note that if the 95% confidence bands include a rate consistent
with normal aging, the observed change may be due to normal aging process rather than
glaucomatous loss.
Normative Data Details Reports
There is measurement variability for the macula parameters which may impact the
normative database color coding. If the true value is near the limit of what the software
uses to determine the normative database color code, then it is possible that the color
code could vary from exam to exam. When at least one parameter is close to a normative
limit, a blue icon button
displays. When your cursor hovers over this icon button, the
tooltip appears as shown below.
cáÖìêÉ=NMJNO==jÉ~ëìêÉãÉåí=s~êá~Äáäáíó=qççäíáé
If you click the blue icon button, you can select Print Preview or Export to DICOM. A Print
Preview screen displays the Normative Data Details Report as shown below. The report can
be printed from the Print Preview screen.
Macula Thickness Normative Data Details Report
The Normative Data Details Report displays the parameters of the analysis in the units
measured and as a percentile from the normative database. In addition, the report displays
the result minus the reproducibility limit, as well as the result plus the reproducibility limit,
and their corresponding percentiles. Each of these measurements are presented with the
appropriate normative database color code.
This report provides the ability to see how close a particular measurement comes to a
normative limit cutoff by checking the actual percentile. In addition, the plus measurement
variability and the minus measurement variability values are also color–coded so the user
can determine if the original normative database designation would cross over to a
different color level when considering the measurement variability. Due to measurement
variability, it is possible that the normative data color coding may change on subsequent
visits without representing a change in the condition of the patient. For more information
on measurement variability, please see
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...