CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Anterior Segment Scans
6-7
Scan Acquisition Controls
Not all posterior scan acquisition controls are available for anterior segment scans. For
anterior segment scans:
• There is no fundus image, and therefore the Auto Focus button and Z controls (left–
right Focus arrows) are not displayed. However, the Focus bar is still displayed,
showing the last focus for the patient.
• FastTrac is not available for anterior segment scans, and therefore the three FastTrac
buttons below the Capture button are not displayed.
• The Optimize button is available only for Anterior Segment Cube 512x128 and Anterior
Segment 5 Line Raster scans.
• The Auto–Enhance button, the Auto–Center button, and manual–center controls are
not available. The OCT display can be centered vertically in the live OCT window by
using the chinrest control buttons or the mouse scroll wheel. However, the shift key +
mouse scroll wheel does not bring the scan into the acquisition window for anterior
scans as it does for posterior scans.
Scan Pattern Adjustments
The scan pattern for anterior scans is displayed on the iris image. The scan pattern cannot
be moved, and scan length is not adjustable. Rotation and line spacing are adjustable for
the Anterior Segment 5 Line Raster scan. Rotation is adjustable for the HD Angle, HD
Cornea, Anterior Chamber, and Wide Angle-to-Angle scans. For more information on
adjusting scan patterns, see
"Scan Pattern Adjustments" on page 6-5
Aligning Scans Corrected for Beam Scanning Geometry and
Corneal Refraction
Anterior Chamber, Wide Angle-to-Angle, HD Cornea, HD Angle, and Pachymetry scans are
corrected to account for beam scanning geometry and refraction on the corneal surfaces.
These corrections are most accurate when acquired corneal scans are centered on the
corneal vertex, which generates a strong central reflection line on the live OCT image.
Typically the corneal vertex is just to the nasal side of the pupil center.
Center Corneal Scans on the Corneal Vertex
NOTE: HD Angle scans are
not
aligned to the corneal vertex.
1. Instruct the patient to fixate on the center of the fixation target, even though it may not
appear to be in focus.
2. Follow the alignment guidelines for each scan to position the scan in the OCT viewport,
making adjustments until there is a strong central reflection line indicating the scan is
centered on the corneal vertex.
Anterior Chamber and Cornea External Lenses
Four anterior segment scans require an external lens:
Anterior Chamber External Lens:
• Anterior Chamber Scan
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...