Appendix
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
A-28
Description of RNFL Parameters
The RNFL Parameters that are compared to normal limits include:
• Average RNFL Thickness around a measurement circle with 3.46 mm diameter.
• Average RNFL Thickness in 4 quadrants distributed evenly around the measurement
circle (Superior, Temporal, Inferior, and Nasal).
• Average RNFL Thickness in 12 clock–hours (30º segments of the measurement circle).
• Thickness in a TSNIT profile around the measurement circle (255 points distributed
evenly around the measurement circle, starting at the temporal most point and
traveling superior – nasal – inferior and back to temporal, or TSNIT).
• RNFL Symmetry: correlation coefficient between left eye TSNIT profile and right eye
TSNIT profile.
Age Coefficient – RNFL Thickness
Analysis of subject demographics determined that expected thickness was dependent upon
age, although the effect was small, and sometimes positive (slightly increasing thickness
with age). Thus age correction is incorporated into the calculated results. Subject ethnicity
was self–reported by the subjects in the population comprising the normative database but
was not used as a variable in constructing the RNFL and macula normative databases.
below displays a scatter plot for average RNFL thickness versus age along
with the fitted regression line.
cáÖìêÉ=^JNR==^îÉê~ÖÉ=okci=qÜáÅâåÉëë=E”ãF=sÉêëìë=^ÖÉ
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...