Appendix
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
C-8
Table C-6: Mean Difference between CIRRUS HD-OCT 5000 and Visante OCT for the
Post-LASIK Group
These tables show that the central cornea as measured using the pachymetry scan has
much better repeatability and agreement with Visante OCT than the central corneal
thickness as measured with the Anterior Chamber scan. The improved performance is likely
due to the fact that the pachymetry measurement averages over a 3mm central area, while
the Anterior Chamber scan depends on subjective placement of the scan directly over the
central cornea for a single measurement. For this reason, we recommend using the
pachymetry scan to get the best possible estimate of central corneal thickness.
NOTE: Measurements made with the CIRRUS HD-OCT Anterior Segment scans should not
be directly compared with Visante OCT measurements.
Scan Type
CIRRUS
Visante
Difference
95% CI
95% LOA
Parameter
N
Mean (SD)
Mean (SD)
Mean (SD)
Mean Difference p-value Mean Difference
Pachymetry
Center
40
465.6 (44.3)
463.7 (44.8)
1.8 (6.1)
-0.1, 3.8
0.064
-10.3, 13.9
Inner Nasal
40
515.8 (39.8)
515.8 (38.9)
0.0 (14.7)
-4.7, 4.7
1.000
-29.4, 29.4
Inner Superior
40
509.5 (39.6)
515.6 (39.4)
-6.2 (13.6)
-10.5, -1.8
0.007
-33.4, 21.1
Inner Inferior
40
500.6 (39.1)
499.2 (39.2)
1.4 (10.8)
-2.1, 4.9
0.418
-20.2, 23.0
Inner Temporal
40
481.7 (39.0)
490.7 (38.5)
-9.0 (9.5)
-12.0, -6.0
<.001
-28.0, 10.0
Outer Nasal
40
584.8 (35.3)
591.9 (36.1)
-7.2 (19.3)
-13.3, -1.0
0.024
-45.7, 31.4
Outer Superior
40
580.7 (35.3)
595.9 (35.6)
-15.3 (18.5)
-21.2, -9.4
<.001
-52.2, 21.6
Outer Inferior
40
562.5 (35.1)
569.1 (37.2)
-6.6 (17.9)
-12.3, -0.9
0.025
-42.5, 29.3
Outer Temporal
40
529.8 (33.6)
552.0 (34.9)
-22.3 (16.4)
-27.5, -17.0
<.001
-55.0, 10.5
N is the number of eyes with measurements.
Difference = CIRRUS - Visante.
ACD of Visante was adjusted by CCT (i.e. ACD = original ACD - CCT/1000).
95% confidence interval (CI) for mean difference is based on t-distribution.
p-value is based on paired t-test.
95% Limits of Agreement (LOA) = mean difference ± 2 x difference SD.
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...