CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Log Files
11-17
2. Select Edit > Move Scan.... The Move Scan dialog opens. To populate the patient list,
either click Search to return all patients, or specify search parameters in the fields
provided. You can also click Advanced Search to search with more parameters.
Select the patient whose file you wish to move the scan data into by left–clicking that
patient name. Click Move to start the move process (you will need to confirm) or click
Cancel or
to close the dialog without moving the scan.
cáÖìêÉ=NNJNQ==jçîÉ=pÅ~å=aá~äçÖ
Log Files
The CIRRUS HD-OCT records the following events and identifies them by date, time, and
User ID:
• Log on/log off
• Display of analysis data
• Creation, modification, deletion of data
• Import/export of data from removable media
• Receipt/transmission of data from/to an external connection (network, for example)
• Remote service activity
The events are automatically recorded in up to 5 audit files of 5 Mb each. When the
maximum limit for files and file size is reached, the CIRRUS overwrites the existing files.
The default folder for the audit log files is
C:\ProgramData\Carl Zeiss Meditec\CIRRUS HD-OCT\Logs.
NOTE: It is highly recommended that you archive audit log files are regularly.
To export audit log files:
1. Log on as administrator.
2. Click Tools > Export Audit Log File.
In the Browse For Folder, browse to the folder you want to export to, and then click Save.
The log is exported as a .zip file with the name: AuditLog_
dd_mm_yyyy_hh_mm
.
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...