Safety
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
1- 6
CAUTION:
When you complete scan acquisition and before you click Finish or ID Patient
in the Acquire Screen, always prompt the patient to sit back and move the head away from
the chinrest. Clicking either of these buttons in the Acquire screen causes the chinrest to
reposition itself beyond the point where the patient’s eye would contact the lens if the
head remained in the chinrest. Failure to observe this warning could result in injury to the
patient.
CAUTION:
The operator should check that the patient is not holding on to the instrument
before or during tests. Although movement of the motorized chinrest is slow, giving plenty
of warning for patients to remove their fingers, there is potential for fingers to be squeezed
and possibly injured.
CAUTION:
(United States) Federal law restricts this device to sale by or on the order of a
licensed healthcare practitioner.
CAUTION:
Do not reconfigure system components on the table, nor add non–system
devices or components to the table, nor replace original system components with
substitutes not approved by Zeiss. Such actions could result in failure of the table height
adjustment mechanism, instability of the table, tipping and damage to the instrument, and
injury to operator and patient.
CAUTION:
Do not use the printer, the instrument, or the optional power table with an
extension cord or a power strip (multiple portable socket outlet).
NOTE: The optional CIRRUS HD-OCT Power Table is safe to use within the patient
environment when the instrument is powered through it, as instructed herein.
Optical Safety
• IEC 60825–1
• EN ISO 15004–2
• Classification: Group 1 Instrument – Per EN ISO 15004–2. Group 1 instruments are
ophthalmic instruments for which no potential light hazard exists.
WARNING: This device contains visual stimuli, including flickering light and flashing
patterns, between 5 and 65 Hz. Medical professionals need to determine whether this
device should be used for patients who may be photosensitive, including those with
epilepsy.
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...