CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Safety
1- 7
CAUTION:
Applicable Phototoxicity Statements (FDA CDRH Ophthalmoscope Guidance
#71): Because prolonged intense light exposure can damage the retina, the use of the
device for ocular examination should not be unnecessarily prolonged. While no acute
optical radiation hazards have been identified for direct or indirect ophthalmoscopes, it is
recommended that the exposure time for the patient’s eye be limited to the minimum time
that is necessary for diagnosis. Infants, aphakes and persons with diseased eyes will be at
greater risk. The risk may also be increased if the person being examined has had any
exposure with the same instrument or any other ophthalmic instrument using a visible light
source during the previous 24 hours. This will apply particularly if the eye has been
exposed to retinal photography. This medical device has no user adjustable intensity
settings for light incident on the retina, nor does it produce UV radiation or short–
wavelength blue light.
CAUTION:
The appliance coupler is the main disconnect device of the instrument.
Position the instrument in such a way to have easy access to disconnect the appliance
coupler in case of an emergency.
CAUTION:
In case of an emergency, disconnect the appliance coupler from the back of
the instrument.
When the power cord is connected to the instrument, the green light on the power switch
will start flashing. When the power switch is pressed, the light will change to solid green
and the entire instrument will be powered.
Networking Safety
WARNING: When networking the CIRRUS HD-OCT, use only network cables with an
unshielded RJ–45 connector. Use of a shielded network cable in the CIRRUS HD-OCT could
result in electrical shock to the patient and/or examiner.
Network Attached Storage Device Safety
WARNING: To directly connect the NAS device to the CIRRUS HD-OCT, use a network patch
cord only with an unshielded RJ–45 connector. Use of a shielded network patch cord will
ground the NAS device through the CIRRUS HD-OCT, which could result in electrical shock
to the patient and/or examiner.
CAUTION:
We strongly recommend you use peripheral devices supplied or approved by
Zeiss, when available, because they will have been tested to work with the instrument. If
you do use a peripheral device that conforms with the requirements in this section but is
not supplied by Zeiss, do not install any unapproved third party software on the instrument.
Installation of any unapproved software, including drivers, could degrade the performance
of the instrument and/or lead to corrupted diagnostic or therapeutic information and may
void the instrument warranty.
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...