Anterior Segment Scans
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
6-6
cáÖìêÉ=SJN==ea=R=iáåÉ=o~ëíÉê=pÅ~å=m~ííÉêå=lå=^ÅèìáêÉ=cìåÇìë=fã~ÖÉ
Enhanced Depth Imaging (EDI)
Enhanced Depth Imaging (EDI) is an optional mode for single and multi–line raster scans
that improves visibility of structures at the bottom of B-scans.
The signal to noise ratio in OCT scans varies across the axial range. The default CIRRUS
setup is such that the best signal is obtained at the top portion of the scan. Enhanced
Depth Imaging allows you to change the acquisition settings for the raster scans so that
the best signal to noise ratio is obtained at the bottom of the B-scan. This allows you to
obtain an HD image that is optimized in the region that is of interest for a particular scan.
To switch between EDI and standard scanning mode, select the EDI checkbox
below the fundus image.
Anterior Segment Scans
Scan Preparation
Instrument Imaging Process
When an anterior segment scan is selected or external lens mounted:
• The LSO illumination of the retina is turned off.
• The internal fixation target is centered. For all anterior segment scans, the patient sees
the green fixation target against a black background. The flashing red lines showing
the scan pattern of the selected scan type are blurry.
• The iris illumination is dimmed by default to avoid causing pupillary constriction.
• The internal lens audibly clicks as it is brought into position.
Drag blue bar left or right
to adjust line length
Drag blue bar up or down
to adjust line spacing
Drag rotation icon to adjust
the angle of rotation
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...