Performance Verification Check
2660021169012 Rev. A 2017-12
CIRRUS HD-OCT User Manual
12-12
In effect, when there is a failure condition, you can clearly see that one or both of the
scan line indicators fail to pass within, even marginally, the white cross–hair in the
center.
cáÖìêÉ=NOJU==bñ~ãéäÉë=çÑ=c~áäìêÉ=`çåÇáíáçåë
NOTE: Evaluation is somewhat subjective. We offer the examples above as guidelines. If
you drag the scan lines, you will observe that there is only a two or three pixel range of
movement while still within the pass condition, and only a one pixel difference between a
marginal pass and a failure. The center cross–hair defines a stringent range of tolerance.
Therefore, you should confirm a failure only if the scan line indicators lie wholly within the
center of the cross–hair pattern.
• To confirm your observation, you should switch back to 0% Transparency.
• If your observation is confirmed initially, you should remove and re–install the test
tool to ensure it is seated properly and then run the check again.
• If the system still fails the test, contact Zeiss customer service. In the U.S., call 800–
341–6968. Outside the U.S., contact your local Zeiss distributor.
If you want to repeat the test, we recommend you first remove and re–install the test tool.
Fail—horizontal line clearly within black
Fail—vertical line clearly within black
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...