CIRRUS HD-OCT User Manual
2660021169012 Rev. A 2017-12
Appendix
A-23
Axial Length and Refractive Error
below shows a plot of Ganglion Cell Analysis Average Thickness versus axial
length of the study eye. It can be seen that GCL + IPL thickness decreases slightly as a
function of axial length. This contributes less than 2% of the total variability of the
Ganglion Cell Analysis parameters.
cáÖìêÉ=^JNQ==^ñá~ä=iÉåÖíÜ=~åÇ=oÉÑê~ÅíáîÉ=bêêçê
Data Analysis
The same analysis model was used as was used for the Macular Thickness normative data
analysis, see
"RNFL and Macula Normative Databases: Diversified" on page
.
A regression model analysis was used to estimate the normative limit of each of the
Ganglion Cell Analysis parameters adjusted by age. The subject's age is considered as a
clinically important factor for the ganglion cell plus inner plexiform layer thickness
measurements.
For each fitted regression model, the residuals were derived for each eye by subtracting the
estimated expected mean reading, ET(age0), from the measured or observed reading,
Obs(age0). In other words, residual = Obs(age0) – ET(age0). The goal was to predict the
100x
th
percentiles (NL, normative limit) of the residuals, so that the 100x
% limit of the
CIRRUS HD-OCT parameter readings could be estimated as follows:
ET(age0) + NL(100x
%) < Obs(age0) (A)
The 1st, 5th, 95th, and 99th percentiles of the residuals were estimated by the empirical
distribution of residual. Then the estimated 1%, 5%, 95% and 99% normal limits of
CIRRUS HD-OCT parameters for a normal subject with an age of age0 were established by
Equation (A). It should be noted that the study site effect was not considered in the
normative limits calculation since the objective was to establish the normative limits for the
general population.
Содержание CIRRUS HD-OCT 500
Страница 1: ...2660021156446 B2660021156446 B CIRRUS HD OCT User Manual Models 500 5000 ...
Страница 32: ...User Documentation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 2 6 ...
Страница 44: ...Software 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 3 12 ...
Страница 58: ...User Login Logout 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 4 14 ...
Страница 72: ...Patient Preparation 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 5 14 ...
Страница 110: ...Tracking and Repeat Scans 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 6 38 ...
Страница 122: ...Criteria for Image Acceptance 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 7 12 ...
Страница 222: ...Overview 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 9 28 ...
Страница 256: ...Log Files 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 11 18 ...
Страница 272: ...Electrical Physical and Environmental 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual 13 4 ...
Страница 292: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 18 cáÖìêÉ JV kçêã íáîÉ a í aÉí áäë oÉéçêí ...
Страница 308: ...Appendix 2660021169012 Rev A 2017 12 CIRRUS HD OCT User Manual A 34 ...
Страница 350: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 I 8 ...
Страница 351: ...CIRRUS HD OCT User Manual 2660021169012 Rev A 2017 12 ...