START-UP
Axioskop 40
Carl Zeiss
Attachments and conversions
Axioskop 40 FL
2-26
B 40-810 e 12/01
2.2.7
Changing the burner for the
HBO 50 fluorescence illuminator
•
Turn off the HBO 50 transformer. Remove the
HBO 50 plug from the transformer socket.
Danger of getting burned!
Allow lamp housing to cool down for
at least 15 minutes.
•
Remove the HBO 50 fluorescence illuminator
from the epi-fluorescence component and place
it on a working surface.
•
Undo clamping screw (2-32/
1
) with the SW 3
ball-headed screwdriver and remove the lamp
housing upwards.
•
Press spring lever (2-32/
4
) down and pull the
heat sink (2-32/
2
) with the burner (2-32/
3
) up
out of the lamp holder (2-32/
5
). Put the heat
sink down, with the clamping screw accessible.
The cable on the heat sink should
not be removed.
•
Undo the clamping screw on the heat sink with
the Allen key and remove the spent burner.
Dispose of the spent burner according to regulations. Observe the manufacturer’s
instructions!
•
Hold the new burner by the labeled metal base and place it into the heat sink, so that the reflective
side of the burner faces down when placed into the lamp holder, or, if both sides are reflective, so
that the electrode labeled "UP" is on top. If there is a side reflective portion of the burner, it must
point to the side in relation to the beam path.
•
Tighten the clamping screw on the heat sink carefully.
Fig. 2-32
Changing the burner
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...