CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axioskop 40
Carl Zeiss
Troubleshooting
Axioskop 40 FL
4-6
B 40-810 e 12/01
Problem
Cause
Remedy
Greater focus differences after
objective change.
The focusing eyepieces are not set
correctly.
Set the focusing eyepieces to the
appropriate ametropia, s. P. 2-6.
The objective is not screwed in to
the stop position.
Screw in the objective to the stop
position.
The tube lens is either not inserted
or inserted superfluously.
Insert the tube lens or remove the
superfluous tube lens.
Interpupillary distance of the
binocular tube is not set correctly.
Set the interpupillary distance
correctly, see P. 2-7.
Left and right fields of view
cannot be combined to one
image.
The focusing eyepieces are not set
correctly.
Set the focusing eyepieces to the
appropriate ametropia, s. P. 2-6.
Eye-fatiguing microscopy.
Interpupillary distance of the
binocular tube is not set correctly.
Set the interpupillary distance
correctly, see P. 2-7.
The focusing eyepieces are not set
correctly.
Set focusing eyepieces to the
appropriate ametropia, s. P. 2-6.
Image brightness is not
acceptable.
Adjust the lamp voltage or insert
the conversion filter.
Binocular tube optically /
mechanically out of alignment.
Check / repair by microscopy
service.
Dirt or dust in the field of view. The condenser is not focused
correctly and front lens 0.9 not
switched correctly.
Focus the condenser and switch
front lens 0.9 on or off correctly,
see P. 3-22 ff.
The aperture diaphragm opening
is too small.
Set the aperture diaphragm
opening as per the 2/3 rule or the
specimen features, see P. 3-22 ff.
Dirt or dust on the optical surfaces
of the objectives, eyepieces,
condensers, filters or specimens.
Clean the optical surfaces of the
affected components,
see P. 4-2.
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...