START-UP
Axioskop 40
Carl Zeiss
Attachments and conversions
Axioskop 40 FL
2-42
B 40-810 e 12/01
2.2.16
Attaching the multi-observation equipment
We recommend that you request Zeiss Microscopy Service to assemble the multi-observation
equipment (refer to Chapter 4), since careful optical adjustment and centering of the fields of
view of the various observation tubes is required.
Should you choose to assemble the equipment yourself, please proceed as follows:
•
Remove binocular tube and intermediate tube (if provided) from the stand (also refer to Section
2.2.10).
•
Insert the base unit of the multi-observation equipment into the stand opening and tighten with
clamping screw (2-48/
1
).
☞
Prior to attachment, the tube lens must be removed from the binocular tube (or phototube) for
the main user! The D-filter provided must be screwed in instead of the tube lens. This brings
the brightness of the image / pointer to the same level as the co-observation tubes.
•
Insert the binocular tube or phototube 30°/23 (2-48/
4
) into the dovetail of the multi-observation
equipment (2-48/
3
), align it and tighten with clamping screw (2-48/
2
).
•
Attach connecting piece L (2-48/
5
) or end connector L (2-48/
7
) to the left side of the multi-
observation equipment, and connecting piece R (2-48/
13
) or end connector R (2-48/
11
) to the right
side. Do not forget to remove the protection caps from the tube ends. Each pin must engage in the
corresponding opening, then tighten the coupling nuts.
•
Attach and clamp two binocular tubes for the multi-observation equipment (2-48/
6
) on each of the
connecting pieces L (2-48/
5
) and R (2-48/
13
).
•
When mechanically aligning the connecting pieces and end connectors, take care to ensure that field
centering, based on the main observer tube (2-48/
4
), is retained. For this purpose, select a
conspicuous specimen area and place it at the point of intersection of the eyepiece reticule (also refer
to Section 1.6). After assembly of the first connecting piece, insert the eyepiece with reticule in the
respective co-observation tube and, in particular, adjust the height of the connecting tube using the
set screw (2-48/
12
) until the selected specimen area is again positioned at the point of intersection of
the eyepiece reticule.
•
Attach and tighten a further connecting piece L (2-48/
5
) or end connector L (2-48/
7
) to the
connecting piece L, and a further connecting piece R (2-48/
13
) or end connector R (2-48/
11
) to
connecting piece R.
•
Attach and clamp two binocular tubes for the multi-observation equipment on each of the end
connectors R and L. Perform height adjustment using the eyepiece reticule.
•
Insert plug (2-48/
9
) of power unit (2-48/
10
, 115 V or 230 V) into the socket (2-48/
8
) on the multi-
observation equipment and tighten the ring to prevent shifting.
•
Connect power unit for light pointer (2-48/
10
) to a power outlet (115 V or 230 V).
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...