OPERATION
Axioskop 40
Carl Zeiss
Axioskop 40 FL operation and function controls
Axioskop 40 FL
3-12
B 40-810 e 12/01
Compartment for filter slider (3-7/5)
−
For 3-position or 6-position filter slider with 18 mm filter diameter.
Centering screws for aperture diaphragm (incident light) (3-7/6)
−
Center the aperture diaphragm with an SW 3 ball-headed screwdriver (also see Section 3.5.7).
Push rod for aperture diaphragm (incident light) (3-7/7)
−
Infinitely variabe setting of the aperture diaphragm diameter (see Section 3.5.7).
−
Push rod pushed in:
fully open
−
Push rod pulled out:
locked
Adjustment guide for lamp adjustment (3-7/8)
−
The adjustment guide is used to adjust and center the HBO 100 and HBO 50 fluorescence illuminators
Light intensity regulator (3-7/9)
−
Regulates the direct-current voltage supply for the halogen illuminator in the range from approx. 3 to
12 V.
On / Off switch with control display (3-7/10)
−
Position 0 = Instrument turned off.
−
Position 1 = Instrument turned on, display glows green when the instrument is turned on.
Focusing drive - rough adjustment (rough drive) (3-7/11)
−
Focusing drive for rough adjustment available on either side of the instrument:
−
1 rough rotation = approx. 5 mm
−
Overall lift: approx. 24 mm
Focusing drive - fine adjustment (fine drive) (3-7/12)
−
Focusing drive for fine adjustment available on either side of the instrument:
−
1 fine rotation = 0.1 rough rotation
Color temperature key 3200 K (3-7/13)
−
With the conversion filter switched on and full lamp voltage (12 V), the color temperature comes to
3200 K. This is required for photomicrography using color films (artificial light).
−
Lever in the front position: conversion filter swiveled out
−
Lever in the back position: conversion filter swiveled in
Drive knob for movement of the mechanical stage in the y direction (3-7/14)
−
Movement of the mechanical stage in the y direction, maximum travel range: 50 mm
−
Adjustable torque (smoothness)
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...