![background image](http://html1.mh-extra.com/html/zeiss/axioskop-40/axioskop-40_operating-manual_3382221068.webp)
START-UP
Axioskop 40
Carl Zeiss
Attachments and conversions
Axioskop 40 FL
2-38
B 40-810 e 12/01
2.2.14
Changing the beam splitter in the FL P&C reflector module
Attachment of filters and the beam splitter requires utmost care to prevent damage to and
contamination of the optical components.
We recommend ordering completely equipped FL
P&C reflector modules, since changing the beam
splitter is quite demanding.
However, should you choose to change the beam
splitter, proceed as follows:
•
Remove the FL P&C reflector module from the
reflector turret (also refer to Section 2.2.12).
•
Undo the two slotted screws (2-44/
1
) with a
screwdriver.
•
Hold both halves of the reflector module
together (
emission
half (2-44/
2
) and
excitation
half (2-44/
3
)), turn in the opposite
direction to the installation position and put it
down.
•
Tip up the
excitation
half of the module
(2-45/
1
), which now is on top, and remove it
from the retaining pins (2-45/
5b
) on the
bottom half of the module (
emission
) (2-45/
4
).
•
Remove the beam splitter (2-45/
2
) and spring
frame (2-45/
3
) from the bottom half of the
module.
•
Remove the old beam splitter and carefully
place the new one onto the spring frame
(2-45/
3
) with the reflective side facing up and
place both parts together into the bottom half
of the module. Take care to ensure that the
side latch of the spring frame is in the
appropriate recess in the bottom half of the
module.
Fig. 2-44
Changing the beam splitter
Fig. 2-45
Changing the beam splitter
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...