CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axioskop 40
Carl Zeiss
Troubleshooting
Axioskop 40 FL
4-4
B 40-810 e 12/01
4.3
Troubleshooting
Problem
Cause
Remedy
Shadows or inhomogeneous
image brightness in the field of
view; the field is not entirely
visible.
The vis/fot push rod on the camera
tube is not in the correct position
(intermediate position).
Move the vis/fot push rod on the
phototube to the correct position
(end position), see Pages 3-16 /
3-17.
The nosepiece with objective is not
in click-stop position.
Switch the nosepiece with
objective to click-stop position.
The condenser is not set correctly.
Set the condenser correctly
(adjustment, centering),
see P. 3-21 ff.
The aperture diaphragm is not set
correctly.
Set the aperture diaphragm
correctly (centering, opening),
see P. 3-21 ff.
The luminous-field diaphragm is
not set correctly.
Set the luminous-field diaphragm
correctly (centering, opening),
see P. 3-21 ff.
The filter is not inserted correctly
in the filter mount.
Insert the filter correctly in the
filter mount.
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...