START-UP
Axioskop 40
Carl Zeiss
Initial start-up
Axioskop 40 FL
2-8
B 40-810 e 12/01
2.1.6
Mounting objectives
•
Move the mechanical stage with stage carrier
to the lower stop position.
•
Remove dust caps (2-7/
2
) from the respective
openings on the nosepiece.
•
Remove objectives (2-7/
4
) from the case and
screw them into the nosepiece (2-7/
1
) starting
with the lowest magnification (in a clockwise
direction).
•
The specimen marker (2-7/
5
) can be inserted in
any nosepiece position instead of an objective.
Place the protection cap onto the specimen
marker during protracted periods when not in
use. See section 2.1.7 for further details.
☞
Take care to ensure the dust protection caps are placed on the nosepiece positions not in use.
Adapter rings (2-7/
3
) can be placed on the objectives before their insertion into the nosepiece to allow
for visual brightness compensation when quickly switching between objectives Type A-Plan, Achroplan
and Plan-Neofluar of different magnification. This makes it unnecessary to constantly adjust the lamp
voltage when switching objectives.
For magnifications up to 20x, the 000000-1104-290 adapter ring is recommended; it contains a neutral
filter with 10% transmission. For magnifications from 20x to <63x, the 000000-1104-289 adapter ring
(with 25% transmission) is recommended. For magnifications from 63x to 100x, the 000000-1104-291
adapter ring (without neutral filter) is recommended. A steady brightness impression is possible only
when the indicated objective types are placed in the nosepiece; however, the impression is also
dependent on the coloring of the specimen.
Since the rings have a height of approximately 5 mm, an adapter ring must be placed on all objective
positions.
☞
The adapter rings are intended only for use with visual applications, since a slight red cast may
occur with CCD or photographic images. In addition, use of the ergo-phototube 20°/23 with
height adjustment (000000-1104-296) in the higher tube settings may slightly darken the
corners of the camera image.
•
To use an adapter ring (2-7/
3
), simply screw it onto the objective and then insert both the objective
and ring into the nosepiece.
Fig. 2-7
Mounting objectives
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...