INSTRUMENT DESCRIPTION
Axioskop 40
Carl Zeiss
Technical data
Axioskop 40 FL
1-14
B 40-810 e 12/01
HBO 50 transformer mbq52ac-z
Operating environment ..........................................................................................................clean room
Protection class ...................................................................................................................................... I
Protection type................................................................................................................................ IP 20
Line voltage can be changed between ................................. 100, 110, 120, 127 VAC and 230, 240 VAC
Line frequency can be changed between ............................................................................ 50 and 60 Hz
Power consumption when HBO 50 is used ........................................................................... max. 350 VA
HBO 100 W (HBO 100 W/2 and HBO 103 W/2) transformer
Operating environment ..........................................................................................................clean room
Protection class ...................................................................................................................................... I
Protection type................................................................................................................................ IP 20
Line voltage ...........................................................................................................100 VAC …240 VAC
Line frequency ..................................................................................................................... 50 … 60 Hz
Power consumption when HBO 103 is used ................................................................................. 260 VA
Fuses in accordance with IEC 127
Axioskop 40 microscope stand ....................................................................................T 1 A/H, 5x20 mm
Axioskop 40 FL microscope stand ................................................................................T 1 A/H, 5x20 mm
HBO 50AC transformer mbq52ac-z ...................................................................... 100 V, 127 V: 2x T 4 A
......................................................................................................................
220 V – 240 V: 2x T 2.5 A
HBO 100 W transformer......................................................................................... T 3.15 A/H, 5x20 mm
Light sources
Halogen lamp ...................................................................................................................... 12 V / 35 W
Adjustment of the light source ......................................................................continuous, 3 to 12 V
Mercury vapor short-arc lamp...................................................................................................... HBO 50
Power consumption for HBO 50 ............................................................................................ 50 W
Mercury vapor short-arc lamp.............................................................................................HBO 103 W/2
Power consumption for HBO 103 W/2 ................................................................................. 100 W
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...