Axioskop 40
START-UP
Axioskop 40 FL
Attachments and conversions
Carl Zeiss
B 40-810 e 12/01
2-23
2.2.6
Changing the HBO 103 W/2 mercury vapor short-arc lamp
Before opening the lamp housing, ensure that there is no connection to the electrical mains.
HBO 103 W/2 lamps may only be changed after they have cooled down. Allow the HBO 103
microscope illuminator to cool down for some 15 minutes to avoid the risk of burns.
The HBO 103 W/2 lamp may only be removed from the packaging and inserted into the
HBO 103 illuminator if a protective mask and safety gloves are worn.
All electrical terminal connections must be made carefully. Pronounced heat during operation
may result in contacts loosening.
After the average operating time of 300 h has
expired, the HBO 103 W/2 mercury vapor short-arc
lamp must be replaced.
The lamp’s operating time can be checked on the
counter (2-28/
1
) of the HBO 100 W transformer.
Follow the operating steps below to change the
HBO 103 W/2 lamp:
•
Disconnect the power plug and the HBO 103
plug from the HBO 100 W transformer.
•
Wear a protective mask and safety gloves.
Fig. 2-28
HBO 100 W transformer
Содержание Axioskop 40
Страница 1: ...Operating Manual Axioskop 40 Axioskop 40 FL Routine microscope...
Страница 14: ...INTRODUCTION Axioskop 40 Carl Zeiss Axioskop 40 FL 0 14 B 40 810 e 12 01...
Страница 30: ...INSTRUMENT DESCRIPTION Axioskop 40 Carl Zeiss Axioskop 40 FL 1 16 B 40 810 e 12 01...
Страница 77: ...Axioskop 40 OPERATION Axioskop 40 FL Carl Zeiss B 40 810 e 12 01 3 3...
Страница 128: ...OPERATION Axioskop 40 Carl Zeiss Axioskop 40 FL 3 54 B 40 810 e 12 01...
Страница 138: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 40 Carl Zeiss Axioskop 40 FL 4 10 B 40 810 e 12 01...
Страница 146: ...APPENDIX Axioskop 40 Carl Zeiss Axioskop 40 FL A 8 B 40 810 e 12 01...