5. SCREENS
1st Edition : Oct. 27, 2023-00
(2) Automatic calibration, simple zero calibration, during automatic validation,
and blowback operation
Operation of
each process
Output hold
Hold extension time
Srart of each process
End of each process
The time set in the " Replacement time " in the flow time setting is extended. See
5.1.2 Auto Calibration Setting, 5.1.3 Auto zero calibration setting, 5.1.5
Setting Automatic Validation, 5.1.8 Blowback setting, for each process.
Calibration execution
(3)Operation at remote hold
Remote hold
Output hold
With input
Without input
(4) Screen display during hold
The lights up at the top of the screen on the measurement screen.
(5) After the start of each process of various calibrations, automatic validation, and hold operation when
blowback is interrupted, output hold for extended hold time is performed even when the process is
interrupted. See 5.4.3 Automatic calibration, 5.4.4 Auto Zero Calibration, 5.4.5 Automatic Validation,
5.6.1 Blowback for each process.)
(6) Other operations at hold
• Instantaneous measurement value display is not held (only output is held)
• When the hold is set to hold setpoint, the O
2
-correction instantaneous value is calculated and held at
the setpoint.
• The range identification contact output does not switch even if range switching is performed while the
unit is on hold.