IM 11G06A01-02EN
5-32
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Setting range
Ref. O
2
concentration: 0 to 19 vol% (default 12 vol%)
O
2
concentration limit: 1 to 20 vol% (default 20 vol%)
Sample Gas Concentration:
None (0.000 vol% default)
(2) Display settings
Configure settings related to the screen display.
l
Items to set
Aiarm notification screen flash:
When set to On, the measurement screen will blink when an
alarm occurs.
Backlight time:
Backlight turns off after a certain period of inactivity and no alarms. 10/30/60 min selectable.
Backlight will be turned on by touching the screen or by an alarm when the backlight is off.
Home screen auto return time
If there is no operation for a certain period of time and a screen other than the measurement
screen is displayed, the display switches to the measurement screen. Select from 10/60
min. However, this function is not enabled in the calibration-related screen, automatic
validation screen, blowback screen, loop check screen, valve check screen, touch panel
adjustment screen, and alarm screen. This function is also disabled when the settings
screen has not been saved (the save icon is displayed in the upper right corner).
(3) Modbus setting
Modbus is available in this product. Modbus communication using
RS485 is available according to the model specification. Please
make settings here according to your hardware configuration,
and read the technical document (TI 11G06A01-01EN) for more
information about Modbus communication.
l
Items to set
Modbus address Modbus slave address.
Baud rate
Data transmission speed;
Select from 9600 bps/38400 bps/115200 bps.
Parity
Select even/odd/none
l
Setting range
Converter address :
1 to 247 (default 1)
CAUTION
The stop bit is 1 bit when parity is even or odd, and 2 bits when parity is none. Setting parity to
even or odd is recommended.