IM 11G06A01-02EN
2-2
2. SPECIFICATIONS
1st Edition : Oct. 27, 2023-00
Contact capacity:
Leakage current 3 mA or less
when OFF
Insulation
Contacts mutual; non-insulating
Internal circuit; transformer
isolation
Function; Remote hold, average
value reset, automatic
calibration start, simple
zero calibration start,
automatic validation
start, remote range
changeover, blowback
contact for ZR802G,
calibration error for
ZR802G
Digital Communications:
RS-485 (Modbus RTU):
115200/38400/9600 bps
Cable length:
Up to 600 m (115200 bps)
Up to 1200 m (38400/9600 bps)
Shield ground
Operating conditions:
Ambient temperature; 0 to 40°C (IR800G)
0 to 45°C (IR810G)
Ambient humidity;
10 to 90%RH
(at 40°C, no condensation)
Storage temperature -10 to +50°C
Storage humidity;
35 to 85%R.H. (no
condensation)
Power Supply voltage:
Voltage rating; 100 to 240 V AC
Allowable range; 85 to 264 V AC
Power Supply Frequency;
Rated frequency; 50/60 Hz
Allowable range; 47 to 63 Hz
Power Consumption
IR800G;
Max. 110 VA
IR810G;
Max. 110 VA
Dimensions (W x D x H):
IR800G;
483 x 492 x 177 mm
IR810G;
412 x 240 x 615 mm
Weight:
IR800G; approx. 16 kg
IR810G; approx. 17 kg
Finish color: silver gray
Enclosure: steel casing, for indoor use
Material of gas-contacting parts:
Gas inlet/outlet; 316SS (stainless steel)
Internal tubing; 304SS, 316SS,
Fluoropolymer
(PTFE, PFA),
PP, PPS,
fluoroelastomer,
calcium fluoride
(CaF
2
)
Gas inlet/outlet:
Rc1/4 or 1/4 NPT internal
thread
Purge gas flow rate: approx. 1 L/min (when
necessary)