IM 11G06A01-02EN
5-23
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Items to set
Response time Response time setting. Set for each Ch (component).
Output limit setting
Setting to limit current output
Hold Hold setting during maintenance, calibration, etc.
On/Off
On/Off setting of hold function
Hold type
Setting output value at hold
Hold set value
Set the output value when the hold type is "setpoint".
Hold (in case of failure) Failure hold setting. Set "On/Off", "Type of hold", and "Hold setting value"
as well as the hold setting.
Current output x
Ch setting assigned to current output pins
(x: 1 to 4)
Output hold
Output hold refers to holding the current output to the previous value or a preset value when the
equipment is under maintenance, in calibration, or in the presence of an error. The equipment
status and the hold types that can be selected at that time are shown in Table 5.1.
Table 5.1
Equipment Status
Selectable
Type of hold
Under maintenance
Under calibration (*1)
During validation
During blowback
An error has occurred.
Off
✓
last value
✓
✓
set value (0% FS to 100% FS)
✓
set value (2.4 mA to 21.6 mA)
✓
*1:
During calibration, manual calibration, automatic calibration, and simple zero calibration are included.