IM 11G06A01-02EN
6-2
6. MAINTENANCE
1st Edition : Oct. 27, 2023-00
6.2
How to replace a fuse
The converter has two fuses. If a fuse blows, replace it as follows.
CAUTION
Before replacing a fuse, be sure to fully investigate the cause of the blown fuse (short circuit, etc.)
and repair it.
(1) Shut off the power supply to the converter to ensure safe replacement.
(2) Remove the fuse from the fuse holder. Using a flat-blade screwdriver that fits into the groove
on the holder cap, rotate the cap 90 degrees counterclockwise. Once this is done, the fuse
can be pulled out with the entire cap.
(3) Install a new fuse in the holder, making sure it has the given rating (5A). Insert the fuse into
the holder with the cap in place, and rotate the cap 90 degrees clockwise while pressing
down with a flat-blade screwdriver.
Fuse used with the product
Maximum rated voltage:
250 V
Maximum rated current:
5 A
Type:
time-lag fuse
Standard:
UL, CSA, VDE certified, Electrical Appliance and Material
Safety Law certified
Part Number
A1513EF
Socket
Fuse
Cap
Flat head screwdriver
Figure 6.1