IM 11G06A01-02EN
2-8
2. SPECIFICATIONS
1st Edition : Oct. 27, 2023-00
(*4) Select the amount you need. The following is for reference only.
/CG1: Wiring only RS-485 and power supply
/CG2: 1-2 Component measurement
/CG3: Three or more component measurements without RS-485 or without external zirconia (ZR22G)/external
oxygen analyzer
/CG4: Other than the above
(*5) Available only when CO is included in the measuring component.
(*6) Available when the O
2
analyzer specification is other than "-N" and one or more of NO, SO
2
, or CO is included in the
measuring component.
(*7) Available when the mount type is "-W" wall mount.
(*8) One pressure regulator valve for the sample gas and one for the reference gas are included. Select this option if input
pressure is not stable.
See <Items specified at order IR800G, IR810G>.
Standard Accessories
Name
Qty
Description
Code Specification
Fuse
2 250 V/5 A delay type 5×20 mm
IEC 60127-2 sheet3
Pre-installed in equipment
Ferrite Cores for Power Cable (*) 1
—
—
Panel mounting bracket
4
—
Mount type "-P"
Cable clip
2
—
none
Screw for fixing cable clip
2
M5, 8 mm length
none
Bolt
4
M8, 35 mm length
"-W"
Washer
8
M8
"-W"
Nut
4
M8
"-W"
*: See Figure 3.5 to attach the ferrite cores.
Spare Parts
Name
Part No.
Qty
Name
Part No.
Qty
C-type snap ring Y9011EV
1 (*1) Filter
K8020PW 1
Plate
K9213FB
1
Snap ring plier K9643ZE
1
*1:
The minimum purchase quantity is 10 per order.
<Items specified at order IR800G, IR810G>
TAGNO. (only if necessary)
You can create TAGNO. (tag number) with alphanumeric characters described in the next table. 16 characters at maximum
can be used.
If you specify TAGNO., it is printed on the stainless name plate/tag label affixed to the instrument.
Symbol (Note)
- Hyphen
_ Underscore
= Equal
+ Plus
/ Slash
: Colon
( Left parenthesis
) Right parenthesis
# Hash
! Exclamation mark
. Period
Number
0, 1, 2, 3, 4, 5, 6, 7, 8, 9
Upper case alphabetics A, B, C, D, E, F, G, H, I, J, K, L, M, N, O, P, Q, R, S, T, U, V, W, X, Y, Z
lower case alphabetics
a, b, c, d, e, f, g, h, i, j, k, l, m, n, o, p, q, r, s, t, u, v, w, x, y, z
(Note): The spaces specified for the TAGNO. are removed. The string is left-squeezed.
n
Measuring gas range
Select the range for the sample gas selected under "Measuring Component". Select "None
(-NN)" for gases not included in the measurement component. This product is free-range.
You can set the range within the selected measurement range. The measurement accuracy
varies when the measurement range is set within the optional range. For details, please refer to
Performance.
For multi-component meters, the measurement ranges for NO/SO
2
/CO/CH
4
measurement cannot
be combined with "-E
�
" and "-P
�
" together.
(Example) Measuring component code - B1 (NO+SO
2
meter)
NO measurement range: -E3 (0-200/2000 ppm), SO
2
measurement range: -E6 (0-500/5000 ppm)
=> Both are -E ranges, so they can be combined.
NO measurement range: -E3 (0-200/2000 ppm), SO
2
measurement range: -P1 (0-2/10 vol%)
=> Not possible due to a mix of -E and -P ranges