IM 11G06A01-02EN
5-53
5. SCREENS
1st Edition : Oct. 27, 2023-00
CAUTION
If a setting confirmation dialog box appears and blowback cannot be started, you have
insufficient contact output setting. See 5.2.3
Contact output setting to assign the necessary
functions.
l
Forced abort
"You can forcibly interrupt the blowback during blowback."
1 Tap the
on the measurement screen during blowback, and the
abort confirmation dialog box will appear. (If an execution password has
been set, the password must be entered before the dialog appears. For
passwords, see 5.3.1 Password.)
2 (When you tap the "Yes" button in the cancel confirmation dialog, the
blowback is canceled, the replacement time is in progress, and
icon
blinks.
3 To stop processing during the replacement time, also touch the
icon that is blinking during the replacement time.
l
Blowback operation
Once the blowback starts, the contact output opens and closes alternately about every 10
seconds for the preset blowback time. After the blowback time has elapsed, the current output is
held in the state set by "hold" until the replacement time elapses.
Blowback start
(Contact input or touch panel)
Contact output
Current output
Blowback interval
Replacement time
Output hold time
10 sec.
ON
10sec.OFF
10 sec.
ON