IM 11G06A01-02EN
5-15
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Example of screen when alarm occurs
When an upper limit alarm occurs, the
icon lights up on Ch
(component).
(In the case of a L-Limit alarm
, HH-Llimit :
, LL-Limit:
are displayed.)
Figure 5.6
NOTE
No alarm judgment is made for 10 minutes after power-on.
5.1.7
Peak alarm setting
During measurement, the peak count alarm turns ON when the number of peaks exceeding the
upper limit of CO measurement concentration exceeds the set number of times. To output the
peak count alarm externally, assign the peak count alarm to any contact output. (For details on
contact output settings, see 5.2.3 Contact output setting.)
This setting screen is displayed only when the option (/A) for peak alarm is specified.