IM 11G06A01-02EN
5-49
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Contact output during automatic validation
The contact output during automatic validation changes as shown in Figure 5.20 and Figure 5.21
according to the gas flow channel setting in the automatic validation setup. (See 5.2.3 Contact
output setting for the functions assigned to contact outputs and 5.1.5 Setting Automatic
for sample switching
for zero calibration
for CH1 span calibration
for CHx span calibration
Validation start
End of validation
Figure 5.20
Gas flow channel setting: Analyzer
End of validation
Validation start
System validation
For zero calib.
For CH1 span calib.
For CHx span calib.
The contact output for "sample switching" remains ON.
Figure 5.21
Gas flow channel setting: Sampling system