IM 11G06A01-02EN
5-46
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Cancel automatic zero calibration
Calibration can be forcibly discontinued during automatic zero calibration.
(1) Tap the
on the measurement screen during automatic zero
calibration, and the confirmation dialog box for cancellation will be
displayed. (If an execution password has been set, you must enter
the password before the dialog appears. For passwords, see 5.3.1
(2) In the dialog box to confirm discontinuation, tap the [Yes] to stop the
calibration. The replacement time will begin, and the
will blink.
(3) If you tap the blinking
during the replacement time, you can also
stop the process during the replacement time.
l
Contact output during automatic zero calibration
The contact output during simple zero calibration shifts as shown in Figure 5.17 and Figure 5.18
according to the gas flow channel setting in the Simple Zero Calibration Setup. (See 5.2.3
Contact output setting for the function assigned to the contact output and 5.1.3 Auto zero
calibration setting for the gas flow channel.)
for sample
switching
for zero
calibration
Calibration
start
End of calibration
Figure 5.17
Gas flow channel setting: Analyzer
for system
calibration
for zero
calibration
Calibration
start
End of calibration
The contact output for "sample switching" remains ON.
Figure 5.18
Gas flow channel setting: Sampling system