IM 11G06A01-02EN
5-7
5. SCREENS
1st Edition : Oct. 27, 2023-00
5.1.3
Auto zero calibration setting
Automatically performs zero point calibration at a set time.
The component to be calibrated is determined by
Auto calibration/Auto zero calibration
setting in 5.1.1 Calibration channel setting.
(1) Tap
on the Home/Main screen > Menu
(2) Tap [Cal/Alarm Setting] > Cal/Alarm settings
(3) Tap [Auto zero cal.setting] > Auto zero calibration setting
(4) Tap each item to set.
l
Items to set
On/Off
On/Off setting of the automatic zero calibration
First start day Set the starting day of the week for the first calibration
First start time [hhmm] Set what time and minute to start the first calibration
Interval
The time interval between the last calibration and the next one.
Interval unit
Set the unit of cycle (hours/day)
Gas Channel Gas flow channel (analyzer/sampling system) setting for automatic zero
calibration
Flow time
Time to replace with calibration gas and time to replace sample gas after
calibration is completed