IM 11G06A01-02EN
5-19
5. SCREENS
1st Edition : Oct. 27, 2023-00
CAUTION
To perform blowback, assign "Blowback" to the contact output. For more information on contact
outputs, see 5.2.3 Contact output setting.
l
Example
On/Off
On
First start day
Sun.
First start time
12:00
Interval
12
Interval unit
hour
Blowback time [min]
3
Blowback time [sec]
20
Replacement
300 sec.
When blowback is performed with the above settings:
:Blowback
in progress
Monday
12
:
00
Monday
00
:
00
Sunday
12
:
00
Cycle
Blowback
3 min.
300 sec
Replacement
300 sec.
l
Setting range
Interval:
1 to 99 hours, or 1 to 40 days (default 7 days)
Blowback time [min]: 0 to 15 minutes (default 10 minutes)
Hysteresis:
60 to 900 seconds (default 180 seconds)
CAUTION
• After the power is turned off (including power failure, etc.) and then turned on again, the next
automatic validation is executed at the time set for the start time, and then at intervals of
cycles.
• Regardless of whether the Blowback is On or Off, the Blowback can be performed from the
screen or by remote start.