IM 11G06A01-02EN
5-51
5. SCREENS
1st Edition : Oct. 27, 2023-00
Touch the combo box
to select component.
The component selection dialog appears, and the component is newly displayed
on the Calibration Log screen.
Select CO.
See the table below for the coefficient name, number of decimal places of the
coefficient, and calibration alarm threshold for each component displayed in the
calibration log. (For component NOx, the coefficient name for zero calibration is
"MAIN," the number of decimal places is 4, and a calibration alarm occurs if the
coefficient is outside the range of -1.0000 to 1.0000). Measured values in the
validation log are displayed with the number of decimal places and units of the
calibration range.
Table 5.4
Each component displayed in the calibration log
Cal.type
Coefficient name
Number of decimal places NOx
SO
2
CO
CO
2
CH
4
O
2
(built-
in)
Zero
MAIN
4
1.0000
-1.0000
1.0000
-1.0000
1.0000
-1.0000
1.0000
-1.0000
1.0000
-1.0000
Coefficient
4
1.0000
0.0000
Span
Coefficient
4
1.666
0.769
1.666
0.769
1.666
0.769
1.666
0.769
1.666
0.769
1.0000
0.0000
For external oxygen meters, only the date, time, and calibration type are recorded, not the coefficient information.
Oxygen meters do not have calibration alarms due to out of coefficient range.
l
Product information
Displays the date and time, Serial No., Software revision, model name and code,
etc. set for this product.
Touch
to scroll through and display all items.