IM 11G06A01-02EN
5-25
5. SCREENS
1st Edition : Oct. 27, 2023-00
Setting of Hold setting value
(1) Tap
on the mA output settings screen/
Tap [Hold setting value].
(2)Tap the item for which you want to change the setting.
Enter the numerical values when the keyboard dialog appears.
l
Setting range
Response Time
0 to 60 s (default 15 s)
Lower limit value
2.4 to 21.6 mA (default 3.8 mA)
Upper limit value
2.4 to 21.6 mA (default 20.5 mA)
Hold setting value (*1) 0 to 100% FS (default 0% FS)
Hold on Fault
2.4 to 21.6 mA (default 3.4 mA)
*1:
The % FS setting is a percentage of the range, and is common to both ranges. For example, if the range is 0-1000ppm and the
hold setting value is set to 10% FS, the output equivalent to 100 ppm is output and held, regardless of the measured value at that
time.
Hold
When "On/Off" of Hold, Hold on Fault is set to "On", the output signal of each channel
(component) is held during the state of each instrument (during calibration validation and
blowback, even during the replacement time after the end of each state). The hold is also turned
On/Off. It is also possible to hold (remote hold) the output signal by external input regardless of
the hold On/Off setting. (See 5.2.4 Contact input setting for details on remote hold.
(1) Operation at manual calibration
When the "Next" button is touched on the calibration content confirmation
screen, gas flows and calibration starts. After that, when the "Enter"
button is tapped on the calibration execution screen, calibration starts and
after the calibration is completed, the replacement time starts.
Tap [ENTER] on the lower right figure
Hold extension time
Calibration
operation
Tap [Next] in the upper right
figure.
Output hold
The time set in the " Replacement time " in the flow time setting is extended.
calibration content confirmation