5-7
IM CW240E
Setting Ranges Using the Direct Keys
5
5.3 Ranges and Number of Digits
●
Active Power Range Configuration
Range Configurations 1 to 10 show the active power ranges corresponding to
the voltage and current ranges. Each value in the table is its full scale value.
TIP
For reactive power or apparent power, the unit for the Range Configuration tables
varies as follows:
• Reactive power (Var)
• Apparent power (VA)
If the VT ratio or CT ratio is set to any value other than "1":
A value obtained by multiplying a value in the Range Configuration tables by the
VT or CT ratio applies.
If the result of "VT ratio
⫻
CT ratio
⫻
1.3" exceeds 9999, the position of the
decimal point is shifted by one digit.
Range Configuration 1 (Using clamp-on probe 96036)
Voltage Range
Wiring
1P2W
30.00 W
75.00 W
150.0 W
300.0 W
1P3W
60.00 W
150.0 W
300.0 W
600.0 W
3P3W
60.00 W
150.0 W
300.0 W
600.0 W
3P4W
90.00 W
225.0 W
450.0 W
900.0 W
1P2W
60.00 W
150.0 W
300.0 W
600.0 W
1P3W
120.0 W
300.0 W
600.0 W
1.200 kW
3P3W
120.0 W
300.0 W
600.0 W
1.200 kW
3P4W
180.0 W
450.0 W
900.0 W
1.800 kW
1P2W
120.0 W
300.0 W
600.0 W
1.200 kW
1P3W
240.0 W
600.0 W
1.200 kW
2.400 kW
3P3W
240.0 W
600.0 W
1.200 kW
2.400 kW
3P4W
360.0 W
900.0 W
1.800 kW
3.600 kW
1P2W
200.0 W
500.0 W
1.000 kW
2.000 kW
1P3W
400.0 W
1.000 kW
2.000 kW
4.000 kW
3P3W
400.0 W
1.000 kW
2.000 kW
4.000 kW
3P4W
600.0 W
1.500 kW
3.000 kW
6.000 kW
96036
Current Range
200.0 mA
500.0 mA
1.000 A
2.000 A
150.0 V
300.0 V
600.0 V
1.000 kV