6-21
IM CW240E
Configuring Settings
6
6.3
General Settings 2/2
The General 2/2 screen allows you to set up the following items:
Setting Item
Description
For the setting up of zero cross filter, see Section 6.8.5, Setting Up ON/OFF of a Zero Cross Filter.
VAR METHOD
SAMPLING METHOD
FREQUENCY SOURCE
AVERAGE TIMES
Wh DISPLAY
INTERVAL Wh DISP.
THD MEASURE METHOD
PA CALC. METHOD
HARM. GRAPH ORDER
Sets whether to use a reactive power meter method.
Sets PLL synchronization or fixed clock (50/60 Hz).
Sets a voltage input level regarded as the reference for frequency
measurement.
Performs moving averaging of measured data.
(Sets the number of averaging times.)
ZERO CROSS FILTER
Sets whether to insert a zero cross filter into the frequency circuit.
Sets the number of digits that can be displayed and the unit of
measurement of electric energy in integration measurements.
Sets the number of digits that can be displayed and the unit of
measurement of interval electric energy in demand measurements.
Sets the calculation method of total harmonics distortion (THD).
Sets the reference for harmonic phase angle.
Sets the order (all orders/odd-number orders) of harmonics to be
displayed on a graph.
NOTE
• Setting items cannot be changed during integration measurements (including
when on standby).
• Unless integral measured data has been cleared, setting items cannot be
changed (with the exception of HARM. GRAPH ORDER). To change a setting
item, press the F5 key (HOLD/Clear) for more than 3 seconds on the Measure
screen to clear integral measured data and then change the setting.