13-2
IM CW240E
13.1 Analog Output (DA Output)
●
Output (4CH)
The CW240 can convert measurement values to direct current voltage signals
and output them.
13.1.1 Analog Output Specifications
Output Voltage :
⫾
1VFS of the rated value of each range
Output Current :
⫾
1 mA maximum
Output Quantity: 4CH
Output Data
: Selected from measurement items
Accuracy
:
⫾
(measurement ac 0.5% F.S.)
Response
: 3 times or less the input signal frequency
(excluding harmonic measurement)
16 times or less the input signal frequency
(harmonic measurement)
Update Cycle
: 1 cycle of the input signal frequency
(excluding harmonic measurement)
16 cycles of the input signal frequency
(harmonic measurement)
●
Response
<Other than harmonic measurement>
DA output is output after the input signal waveform.
The delay (response) is 3 times or less a single cycle of the input signal
frequency (3 cycles or less).
<Harmonic measurement>
DA output is output after the input signal waveform.
The delay (response) is 16 cycles or less of the input signal frequency.
Chapter 13 Using Analog Input/Output (Optional)