13-6
IM CW240E
13.2 Analog Input
13.2.1 Analog Input
●
Input (2CH)
The CW240 can measure a direct current voltage signal.
Electric power measurement data and environmental measurement data can be
managed collectively by measuring the output signals of thermometers and
illuminometers.
13.2.2 Analog Input Specifications
Input Range
: 3 ranges, 100 mV/1 V/5V
Input Resistance : Approx. 100 k
⍀
Accuracy
:
⫾
0.5% of F.S.
Resolution
: Po 11 bits
Sample Rate
: Approx. 20 mS