:SYSTem:COMMunicate:GPIB?
Function
Queries all the GP-IB communication settings.
Syntax
:SYSTem:COMMunicate:GPIB?
:SYSTem:COMMunicate:GPIB:ADDRess
Function
Sets or queries the GP-IB address.
Syntax
:SYSTem:COMMunicate:GPIB:
ADDRess <NRf>
:SYSTem:COMMunicate:GPIB:ADDRess?
<NRf> = 0 to 30
Example
:SYSTEM:COMMUNICATE:GPIB:ADDRESS 1
:SYSTEM:COMMUNICATE:GPIB:ADDRESS?
-> :SYSTEM:COMMUNICATE:GPIB:ADDRESS 1
:SYSTem:COMMunicate:GPIB:TYPE
Function
Sets or queries the GP-IB command type.
Syntax
:SYSTem:COMMunicate:GPIB:
TYPE {NORMal|COMPatible}
:SYSTem:COMMunicate:GPIB:TYPE?
NORMal:
Normal commands that comply with
IEEE488.2-1992
COMPatible: Model2560 compatible commands
Example
:SYSTEM:COMMUNICATE:GPIB:
TYPE COMPATIBLE
:SYSTEM:COMMUNICATE:GPIB:TYPE?
-> :SYSTEM:COMMUNICATE:GPIB:
TYPE COMPATIBLE
:SYSTem:COMMunicate:USB:FUNCtion
Function
Sets or queries the USB feature (USB-TMC/
storage).
Syntax
:SYSTem:COMMunicate:USB:
FUNCtion {TMC|STORage}
:SYSTem:COMMunicate:USB:FUNCtion?
TMC: USB-TMC command control feature
STORage: Storage feature
Example
:SYSTEM:COMMUNICATE:USB:
FUNCTION STORAGE
:SYSTEM:COMMUNICATE:USB:FUNCTION?
-> :SYSTEM:COMMUNICATE:USB:
FUNCTION STORAGE
:SYSTem:REBoot
Function
Restarts the 2560A when the specified time
elapses.
Syntax
:SYSTem:REBoot <
Time
>
<Time> = 0 (S) to 60 (S)
Example
:SYSTEM:REBOOT 3
:SYSTem:RESolution
Function
Sets or queries the resolution mode.
Syntax
:SYSTem:RESolution {NORMal|HIGH}
:SYSTem:RESolution?
NORMal: Normal resolution (5.5 digit) mode
HIGH: High resolution (6.5 digit) mode
Example
:SYSTEM:RESOLUTION HIGH
:SYSTEM:RESOLUTION?
-> :SYSTEM:RESOLUTION HIGH
:SYSTem:RJC:RESister
Function
Sets or queries the resistance of the sensor to be
used for the external reference junction for 0°C.
Syntax
:SYSTem:RJC:RESister <
Resistance
>
:SYSTem:RJC:RESister?
Example
:SYSTEM:RJC:RESISTER 100.00
:SYSTEM:RJC:RESISTER?
-> :SYSTEM:RJC:RESISTER 100.00
:SYSTem:TEMPerature:MONitor
Function
Sets or queries the output value display mode for
temperature sourcing.
Syntax
:SYSTem:TEMPerature:
MONitor {DEGree|VOHM}
:SYSTem:TEMPerature:MONitor?
DEGRee: Temperature display mode
VOHM: Voltage or resistance display mode
Example
:SYSTEM:TEMPERATURE:MONITOR DEGREE
:SYSTEM:TEMPERATURE:MONITOR?
-> :SYSTEM:TEMPERATURE:MONITOR DEGREE
Description If the source function is set to TCouple or RTD
with the :SOURce:FUNCtion command, setting
VOHM with this :SYSTem:TEMPerature:MONitor
command causes the output value display to be
as follows.
TCouple: Voltage display (mV)
RTD: Resistance display (Ω)
:SYSTem:UNIT
Function
Sets or queries the temperature unit.
Syntax
:SYSTem:UNIT {C|CEL|F|FAR}
C, CEL:°C (Celsius)
F, FAR: °F (Fahrenheit)
Example
:SYSTEM:UNIT C
:SYSTEM:UNIT? -> :SYSTEM:UNIT C
Description On models with the -UC suffix code, this
command is invalid. The unit is fixed at °C
(Celsius). On models with the -UF suffix code,
you can set the unit to °C (Celsius) or °F
(Fahrenheit).