6-17
IM 2560A-01EN
Temperature-Sensor-Simulation Thermoelectromotive-Force and Resistance Sourcing
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
4.
Turn a
main setting
dial to select RJC.R0.
• The deviation display shows the difference between the currently set 100.00 Ω and R0.
• The output display shows the current R0 resistance.
Main setting dials (use any)
4. Select RJC.R0.
Difference between 100.00 Ω and R0
Current R0
5.
Turn the
deviation
dials to set the difference between 100.00 Ω and R0.
• When you change the setting, all PRESET indicators turn on.
• If you decide not to change the setting, step 6 is not necessary.
6.
Flip the
PRESET
switch down.
The setting is confirmed. All the PRESET indicators turn off.
Deviation dials
5. Set the difference between
100.00 Ω and R0.
PRESET switch
6. Confirm the settings.
PRESET indicator
6.6 Setting Reference Junction Compensation