App-13
IM 2560A-01EN
Appendix
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
Appendix 7 List of Default Settings and Settings
That Are Saved
Setup items
Factory Default
Value
Initialized When Settings
Are Initialized or with a
*RST Command?
Yes: Initialized
No: Not initialized
Value Saved When
Powered Off?
Yes: Saved
No: Not saved
Output
OFF
Yes
No
Range
1V
Yes
Yes
Temperature type
Thermocouple K
Yes
Yes
Main setting (MAX)
Main setting (MIN)
Voltage, current
0.00000
Yes
No
The decimal place
depends on the range
setting.
Temperature
0.0
Yes
No
Resistance
0.00
Yes
No
Voltage, current polarity
+
Yes
No
Temperature polarity
+
Yes
No
Scale selection
MAX
Yes
No
Output divider
Numerator n
10
Yes
No
Denominator m
10
Yes
Yes
Deviation
0.00%
Yes
No
Preset
0%
Yes
No
Sweep
HOLD
Yes
No
Sweep time
Approx. 16 s
Yes
No
Resolution mode
NORM (5.5 digits)
Yes
Yes
Reference Junction
Compensation
(RJC)
Mode
AUTO (auto)
Yes
Yes
Manual input compensation
value
0.00°C
Yes
Yes
Difference between the
resistance for when the
external RJ sensor is at 0°C
and 100.00 Ω (RJC.R0)
0.00Ω
No
Yes
User definitions
Thermocouple
None
No
Yes
RTD
None
No
Yes
Temperature output display
°
Yes
Yes
Temperature unit (only on models with the -UF
suffix code)
°C
No
Yes
Beep sound
ON
Yes
Yes
GP-IB Interface
addresses
1
No
Yes
Command type (TYPE)
NORMAL
No
Yes
Ethernet
interface (LAN)
DHCP
ON
No
Yes
IP address
192.168.0.1
No
Yes
Subnet mask
255.255.255.0
No
Yes
Default gateway
0.0.0.0
No
Yes
USB interface
TMC
No
Yes