6-16
IM 2560A-01EN
Setting the Reference Junction Temperature When the RJC Mode Is MAN (Manual
input)
4.
If you selected MAN (manual input), turn the right-most
main setting
dial to set the reference
junction temperature.
A main setting dial other than the right-most dial
4. Set the reference junction temperature for manual input.
Setting Resistance R0 of an External RJ Sensor
If you want to use the measured temperature of an external RJ sensor for the reference junction
temperature, set the resistance for when the RJ sensor is at 0°C.
1.
Turn the
RANGE
dial to select SETUP.
The SETUP menu appears.
RANGE dial
1. Select SETUP.
The SETUP indicator turns on.
2.
Turn a
main setting
dial to select TEMP.
“NEXT” appears on the deviation display. All the PRESET indicators turn on.
3.
Flip the
PRESET
switch down.
The TEMP menu appears. All the PRESET indicators turn off.
PRESET switch
3. Display the TEMP menu.
Main setting dials (use any)
2. Select TEMP.
PRESET indicator
6.6 Setting Reference Junction Compensation