16-1
IM 2560A-01EN
Specifications
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
Chapter 16 Specifications
16.1 Source Section
DC Voltage and DC Current Source
DC Voltage Source
Range
Source Range
1
Setting Resolution
2
Maximum Output
Current
3
Output Resistance
4
Maximum Capacitive
Load
100 mV
± 122.400 mV
1 µV
12 mA or higher
6
mΩ
or lower
10 µF
1 V
± 1.22400 V
10 µV
Approx. 120 mA
6
mΩ
or lower
10 µF
10 V
± 12.2400 V
100 µV
Approx. 120 mA
6
mΩ
or lower
10 µF
100 V
± 122.400 V
1 mV
Approx. 30 mA
30
mΩ
or lower
1 µF
1000 V
± 1224.00 V
10 mV
Approx. 10 mA
1000
mΩ
or lower
0.1 µF
Range
Display Resolution
Stability (1h)
5
±(ppm of s V)
Accuracy (±(ppm of s V))
6, 7, 8, 9
Temperature
Coefficient
±(ppm of s V)/°C
180 days
1 year
100 mV
1 µV
20 +
3 µ
40 +
4 µ
60 +
4 µ
5 + 0.3 µ
1 V
10 µV
5 +
5 µ
40 + 10 µ
55 + 15 µ
3 +
1 µ
10 V
100 µV
5 + 50 µ
40 + 100 µ
55 + 150 µ
3 + 10 µ
100 V
1 mV
5 + 500 µ
40 +
1 m
55 + 1.5 m
3 + 100 µ
1000 V
10 mV
5 +
5 m
10
40 + 10 m
10
55 + 15 m
10
3 +
1 m
DC Current Source
Range
Source Range
1
Setting Resolution
2
Maximum Output
Voltage
3
Output Resistance
4
Maximum Inductive
Load
100 µA
± 122.400 µA
1 nA
Approx. 30 V
100
MΩ
or higher
1 mH
1 mA
± 1.22400 mA
10 nA
Approx. 30 V
100
MΩ
or higher
1 mH
10 mA
± 12.2400 mA
100 nA
Approx. 30 V
100
MΩ
or higher
1 mH
100 mA
± 122.400 mA
1 µA
Approx. 30 V
10
MΩ
or higher
1 mH
1 A
± 1.22400 A
10 µA
Approx. 10 V
1
MΩ
or higher
1 mH
10 A
± 12.2400 A
100 µA
Approx. 2 V
10
kΩ
or higher
1 mH
30 A
0 to +36.720 A
1 mA
Approx. 1.5 V
5
kΩ
or higher
1 mH
Range
Display Resolution
Stability (1h)
5
±(ppm of s A)
Accuracy (±(ppm of s V))
7, 8, 9
Temperature
Coefficient
±(ppm of s A)/°C
180 days
1 year
100 µA
1 nA
50 +
5 n
100 + 12 n
150 + 20 n
10 + 0.5 n
1 mA
10 nA
5 + 15 n
50 + 20 n
70 + 30 n
3 + 1.5 n
10 mA
100 nA
5 + 150 n
50 + 200 n
70 + 300 n
5 + 15 n
100 mA
11
1 µA
10 + 1.5 µ
70 +
2 µ
90 +
3 µ
10 + 150 n
1 A
10 µA
25 + 25 µ
250 + 50 µ
350 + 70 µ
15 +
6 µ
10 A
100 µA
50 + 500 µ
300 +
1 m
380 + 1.2 m
30 + 60 µ
30 A
100 µA
70 + 1.2 m
450 + 1.5 m
540 + 1.8 m
30 + 300 µ
1 If you set the main setting (MAX) to 120% of the range, the main setting (MIN) to 0% of the range, the divider value n/m to 1,
and the deviation to –2%, up to 122.4% of the range can be generated.
2 The resolution in high resolution mode is 1 ppm of range. But, for the 30 A range, it is 100 µA.
3 Can be output when the source range is between 1% and 120% of the range.
4 Using measurement leads B8506ZK, 758917, or 758933, excluding measurement lead aging, and excluding the measurement
lead internal resistance.
5 Indicates the 1-hour stability under standard operating conditions with the ambient temperature in the range of 23°C ± 1°C
after 1 hour has elapsed with the voltage or current source output turned on.
6 Excludes the voltage drop due to the output resistance.
7 Specifications under standard operating conditions.
8 If the ambient temperature is between 5°C to 20°C or 26°C to 40°C, add the temperature coefficient compensation to the
accuracy.
9 For source ranges that exceeds 120% of each range, add 500 ppm of range.
10 For source ranges that exceeds 100 V, add 12ppm×(source voltage/1000)
2
of range.
11 Only for the 100 mA range, the specifications are defined for sinking current up to 30 mA. Immediately after sinking current
that exceeds 30 mA, errors exceeding the accuracy specifications may occur due to the increased temperature inside the
2560A.