13-7
IM 2560A-01EN
13-7
Commands
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
13.5 OUTPut Group
The commands in this group deal with the output on/off setting and polarity setting. The commands correspond to the
front panel OUTPUT and POLARITY switches.
:OUTPut[:STATe]
Function
Sets or queries the output’s on/off state.
Syntax
:OUTPut[:STATe] <Boolean>
:OUTPut[:STATe]?
ON|1: Enabled
OFF|0: Disabled
Example
:OUTPUT:STATE ON
:OUTPUT:STATE? -> :OUTPUT 1
Description
If the source function is RJTemp and you set
the output’s on/off state, a “221: Setting conflict”
error will occur. Set the source function with the
:SOURce:FUNCtion command.
:OUTPut:POLarity[:AUTO]
Function
Sets or queries the polarity of the main setting
for main setting (MAX) or main setting (MIN),
whichever is selected.
Syntax
:OUTPut:POLarity[:AUTO] {NORMal|
INVerted}
:OUTPut:POLarity[:AUTO]?
NORMal: Positive polarity
INVerted: Negative polarity
Example
:OUTPUT:POLARITY:AUTO NORMAL
:OUTPUT:POLARITY:AUTO?
-> :OUTPUT:POLARITY:AUTO NORMAL
Description • If the source function is RJTemp and you set
the polarity, a “221: Setting conflict” error will
occur.
NORMal is returned as a response
to a query. Set the source function with the
:SOURce:FUNCtion command.
• If the output setting goes outside the selectable
range of the source function range when you
change the polarity, the output will turn off. In
this condition, the output cannot be turned on.
:OUTPut:POLarity:MAXimum
Function
Sets or queries the polarity of main setting (MAX).
Syntax
:OUTPut:POLarity:MAXimum {NORMal|
INVerted}
NORMal: Positive polarity
INVerted: Negative polarity
Example
:OUTPUT:POLARITY:MAXimum NORMal
:OUTPUT:POLARITY:MAXimum?
-> :OUTPUT:POLARITY:MAXimum NORMAL
Description If the source function is RJTEMP and you set the
polarity, a “221: Setting conflict” error will occur.
NORMal is returned as a response to a query.
:OUTPut:POLarity:MINimum
Function
Sets or queries the polarity of main setting (MIN).
Syntax
:OUTPut:POLarity:MINimum {NORMal|
INVerted}
NORMal: Positive polarity
INVerted: Negative polarity
Example
:OUTPUT:POLARITY:MINimum NORMal
:OUTPUT:POLARITY:MINimum?
-> :OUTPUT:POLARITY:MINimum NORMAL
Description If the source function is RJTEMP and you set the
polarity, a “221: Setting conflict” error will occur.
NORMal is returned as a response to a query.