16-7
IM 2560A-01EN
Specifications
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
Item
Specifications
User definition
Thermocouple Thermoelectromotive force corresponding to the set temperature can be sourced
according to the user-defined polynomial.
RTD
Resistance corresponding to the set temperature can be sourced according to the
user-defined polynomial.
High resolution mode
The main setting of the DC voltage or DC current can be set at 6.5 digit resolution (but 5.5 digits
for the 30 A range). It cannot be applied to resistance or temperature. In high resolution mode, the
output divider, deviation, preset, sweep, and scaling cannot be used.
Saving of setup parameters Saves the setup parameters immediately before (approx. 1 s before) the power switch is turned off.
The next time the power is turned on, the 2560A starts with those settings.
Error log
Logs the error codes of the errors that the 2560A detects. The error log can also be displayed.
Self-test
Performs a self-test on its condition.
Initializing the Settings
Resets the 2560A to its default settings.
Displaying the Product
Information
Displays the firmware version, serial number, and the like.
Communication
Communication commands can be used to remotely control the 2560A. Select the communication
interface from GP-IB, Ethernet, and USB.
Instrument protection
Trip function based on overvoltage, overcurrent, oscillation, and internal heating detection are
available.
16.3 External Input
Item
Specifications
External RJ sensor input
terminal
Supports three-wire and four-wire systems
Input connector
Connect the supplied terminal plug A2119JT.
The input terminals of the terminal plug A2119JT are screw type.
Connect an RJ sensor 257875 or a general RTD Pt100 to the terminal plug A2119JT.
Excitation current
Approx. 1 mA