4-4
IM 2560A-01EN
Deviation Dials
Turn the deviation dials to select or set the settings (options or values) of each SETUP menu.
• For options
You can use either deviation dial 1 or 2 to select the same items.
• For values
Turn the deviation dials to set the value. Carrying over and borrowing occurs automatically within
the selectable range.
• Deviation dial 1
Enters the second digit from the right when facing the 2560A panel. Turning deviation dial 1
typically changes the value by 2. If changing the value by 2 causes it to exceed the upper limit or
go below the lower limit of the selectable range, it is set to the upper or lower limit.
• Deviation dial 2
Enters the first digit from the right when facing the 2560A panel. Turning deviation dial 2 changes
the value by 1.
PRESET Switch
Use this switch to confirm the setting that you have changed or cancel the change and return to the
previous menu level.
• Flipping the PRESET switch down confirms the displayed setting.
• If “NEXT” is displayed in setup parameter 1, the screen advances to the next level menu.
• If you change the current setting with the deviation dials, all PRESET indicators turn on. Flipping
the PRESET switch down in this condition turns off the PRESET indicator and confirms the
updated setting.
• If you flip the PRESET switch down on an execution menu, you cannot return to the previous
condition.
• Flipping the PRESET switch up returns you to the condition before changing the settings.
• If there is a previous level menu, “NEXT” is displayed in setup parameter 1, and the screen
returns to the previous level menu.
• If you change the current setting with the deviation dials and all the PRESET indicators are
turned on, flipping the PRESET switch up will not change the display. The settings will not be
changed.
Note
If you clear the error log, initialize the settings, and so forth, you cannot undo it. Confirm that it is okay to
perform this operation before execution.
4.2 How to Use the SETUP Menu